Steam Generator
Made of heavy stainless steel sheet. Heavy ring mounted in front of the boiler with folding thick stainless steel plate fitted with heating elements and low water level cutoff device to protect the former from burning out dry. Front folding plate system provides for easy cleaning of the deposited scale on the elements and the walls for long life and efficiency. Fitted with water gauge glass for water level indication, water inlet and outlet valves.
Pressure Control
The pressure control device is incorporated in all electrically operated sterilizers. It economizes the power consumption and reduces the frequent opening of the steam release valve and prevents the release of steam in the room, it cuts off the power supply to heaters when the set pressure is achieved an re-energises the heating when the set pressure is achieved and re-energises the heating elements when pressure falls below the set point.
Heating Elements
Flanged type/ immersion heating element are made of high grade material Operating Temp. & Pressure Sterilizing temp : 121" Sterilizing pressure : 1.2o 1.5 kg/cm2 (15 psi to 22 psi) Power Requirement : Suitable to operate on 440V volts, 3ph,50Hz/Ac supply
Size/Capacity of Autoclave:
Item Code |
Size |
Capacity |
1041 A |
16" X 24" (400 X 600 mm) |
80 Ltr |
1041 B |
18" X 36" (450 X 900 mm) |
150 Ltr |
1041 C |
20" X 36" (500 X 900 mm) |
185 Ltr |
1041 D |
24" X 48" (600 X 1200 mm) |
365 Ltr |
|